Request a price
In horizontal flow cabinets, the sterile air flows outward from the entire surface of the back wall of the workspace, then passes through the interior of the cabinet in the direction of the operator – protecting the samples from external contamination – and exits the cabinet from the front.
BIOBASE produces two types of horizontal cabinets: 570 and 650 mm internal height. The cabin with a lower interior height is more equipped than the 650 mm height, as it also has a motorized window, its UV lamp has a timer, and a gas valve is also installed in this cabin. The cabinets are produced in 4 width sizes: 980 and 1,380 mm internal width (with an internal height of 650 mm) and 1,200 and 1,700 mm internal width.
The sterile air (99.999% = H14 purity) is produced by a DOP-tested HEPA filter according to EN1822. The washable pre-filter increases the life of the main filter. If the air speed is not adequate (the filter is clogged), the alarm function is activated.
The material of the cabinets is steel plate protected with antibacterial paint. The safety glass windows on the side ensure adequate visibility, and valves for the introduction of gas,CO2, and vacuum can be mounted on them if necessary.
The work surface is an easy-to-clean stainless steel plate. The 720 mm working height cabinets have the lighting, the UV lamp, the gas valve (the larger ones also have the water tap) and the water-free electric plug. The cabinets with a working height of 600 mm - apart from the pins - come with all the above-mentioned equipment.
All BIOBASE cabinets are delivered with a rollable / leveling stand. The devices are characterized by a low noise level (<60dBA).
BIOBASE cabinets are tested to BS EN ISO 14664-1 Class 5.
Specifications: | BBS-H1100 | BBS-H1300 | BBS-H1500 | BBS-H1800 |
---|---|---|---|---|
Consumption: | 200 W | 400 W | 300 W | 400 W |
Window movement: | manual | motorized | manual | motorized |
UV lamp: | 1 x 20 W | 1 x 30 W | 1 x 30 W | 2 x 20 W |
Lighting (LED): | 1 x 12 W | 1 x 12 W | 1 x 16 W | 2 x 16 W |
Air flow velocity: | 0,3 -0,45 m/s | 0,3 -0,5 m/s | 0,3 -0,45 m/s | 0,3 -0,45 m/s |
Noise: | ≤ 65 dB(A) | ≤ 65 dB(A) | ≤ 65 dB(A) | ≤ 65 dB(A) |
Workplace height: | 750 mm | 750 mm | 750 mm | 750 mm |
Workplace internal height: | 600 mm | 570 mm | 600 mm | 570 mm |
Internal dimensions: | 1000 x 500 x 600 mm | 1200 x 500 x 570 mm | 1400 x 500 x 600 mm | 1700 x 500 x 800 mm |
External dimensions: | 1100 x 808 x 1690 mm | 1300 x 820 x 2040 mm | 1500 x 808 x 1690 mm | 1800 x 808 x 1890 mm |
Weight: | 175 kg | 254 kg | 216 kg | 255 kg |
Nüve laminar air flow cabinets in two size: 90 and 120 cm width workplace, 885 and 1.195 mm internal width.
More informationsSupport stands, UV lamps, vacuum and gas valves.
More informationsBIOBASE vertical flow laminar booths, five types - a total of nine booths - in various sizes/designs.
More informationsPCR cabinet manufactured by BIOBASE, fits in a 100 cm wide space, can be moved on a rolling stand.
More informationsBIOBASE CLASS I biological safety cabinets, 3 benchtop types, 1 free-standing type.
More informationsCLASS II B2 type, 100% exhaust biological safety cabinets in four width sizes (up to 90 - 160 cm internal width), motorized windows, remote control, external pressure fan, +4 meter pipe. Manufacturer: BIOBASE
More informations